The research team from the Department of Radiation Oncology at Kangbuk Samsung Hospital announced on July 31 that they have successfully verified the geometric precision and respiratory tracking performance of a next-generation surface-guided radiation therapy system.


During radiation therapy for breast cancer patients, deep inspiration breath-hold (DIBH) is commonly used to minimize radiation exposure to the heart and lungs by instructing patients to take a deep breath and then hold it briefly. Traditional methods require either tattooing small marks on the patient's body to align their position accurately for each treatment session, or attaching special sensors to the body to draw reference lines and monitor respiratory motion, resulting in patient discomfort.

Professors Junbok Lee (left) and Sucheol Han (right), Department of Radiation Oncology, Kangbuk Samsung Hospital. Kangbuk Samsung Hospital

Professors Junbok Lee (left) and Sucheol Han (right), Department of Radiation Oncology, Kangbuk Samsung Hospital. Kangbuk Samsung Hospital

View original image

One solution to reduce such inconvenience is the next-generation surface-guided radiation therapy system. This technology does not require any marks or sensors to be placed on the patient's body. Instead, it utilizes a built-in 3D stereoscopic camera and thermal imaging to monitor the patient’s skin surface and movements in real time, allowing for precise tracking of their position.


The research team, led by Professors Junbok Lee, Sucheol Han, Heerim Nam, Jindong Jo, and Hyebin Lee, conducted a three-phase experiment using a specialized phantom model that accurately replicates the human body and respiratory motions to determine whether the next-generation surface-guided radiation therapy device functions accurately during breath-hold treatment.


Reflecting the actual treatment process, the experiment was conducted in three phases: (1) verifying the positional reproducibility of the three-dimensional surface imaging system, (2) comparing positional accuracy between the new system and conventional three-dimensional image-guided radiation therapy systems using anatomical phantom models, and (3) assessing real-time respiratory tracking performance using a phantom that simulates four-dimensional dynamic motion.


As a result, the next-generation surface-guided radiation therapy system showed high reproducibility in positional evaluation, with deviations of less than 0.3 mm in the left-right and vertical directions and 0.7 mm in the longitudinal direction. Furthermore, when compared with conventional image-guided radiation therapy systems, the positional error was within 1 mm, demonstrating a high level of positional accuracy equivalent to existing methods.


Notably, in the evaluation with the dynamic phantom replicating actual respiratory motion, the respiratory signals measured by the surface-guided radiation therapy system were almost identical to those obtained by the conventional sensor-based equipment. The amplitude difference of respiratory movement was also found to be less than 0.3 mm. This confirmed the system's ability to precisely monitor the patient's respiratory status during deep inspiration breath-hold treatments.


Professor Junbok Lee from the Department of Radiation Oncology at Kangbuk Samsung Hospital stated, "This study demonstrates that the next-generation surface-guided radiation therapy system not only significantly improves patient convenience but also offers a level of precision equal to proven conventional radiation therapy techniques," adding, "We expect it will greatly enhance the convenience and safety of breast cancer treatment going forward."


Additionally, Professor Heerim Nam from the Department of Radiation Oncology at Kangbuk Samsung Hospital shared future plans, saying, "Based on the results of this study, we will explore the potential for applying this technology to organs such as the lungs or liver, where respiratory-controlled radiation therapy is needed."



Meanwhile, the results of this study were published in the latest issue of the international journal, Radiation Physics and Chemistry.